• CL-2000i Chemiluminescence immunoassay system

    Enhanced Reaction System · Non-touch vortexer mixing for reaction solution without any contamination · Precisely controlled temperature for magnetic separation, incubation & reaction by maintenance-free · · direct-solid heating technology · Reliable clinical performance with enhanced PMT chemiluminescent detection system Efficient Washing System · Interior & exterior probe washing with high-pressure pre-heated water · Programmed enhanced washing with detergent for reagent and sample probes · 4-phase washing for magnetic separation unit with high efficiency Stable Reagent Cartridges Utility · 2°C~8°C constant cooling compartment for reagents cartridges · Membrane-sealed liquid reagent ready-to-use · Intelligent magnetic particles mixing by reagent carousel rotation Intelligent Indication for Daily Operation · Real-time indication of cuvettes, waste, and wash buffer status · Real-time monitoring for substrates consumption · Easy operation for replacing cuvettes, wash buffer and placing new substrates

General Information
Throughput:
Up to 240 tests/hour
Measurement Principle:
Micron superparamagnetic particles platform 
with alkaline phosphatase (ALP) labeled 
reagents and AMPPD substrate
Sample Assembly
Sample Feeder:
Rack system with Sample Delivery Module 
Up to 300 onboard samples with 30 racks
Sample Probe:
Steel probe, liquid level detection, clot detection, 
horizontal and vertical collision protection
Sample Volume:
10µl~200µl, with increment of 1µL
Sample Dilution:
Automatic dilution with 1:2 ~ 1:40
Reagent Assembly
Reagent Disk:
36 reagent positions, continuous loading, 
real-time mixing for magnetic particles
Cooling System:
2°C~8°C
Reagent Barcode:
Automatic scanning
Reagent Probe:
Steel probe, liquid level detection, bubble detection, 
horizontal and vertical collision protection
Reagent Volume:
20µL~135µL, with increment of 1µL
Substrate System
Substrate Bottle: 
500 tests per bottle, bottom piercing
Substrate Loading: 
Capable of continuous loading during testing
Substrate Heating: 
Constant heating
Aspiration:
Volume 200µL
Magnetic Separation Unit
Separation Assembly: 
4-phase magnetic separation
Temperature: 
Accuracy: 37±0.3°C, fluctuation: ±0.1°C
Measurement and Reaction System
Detection Mode:
Photon counting
PMT Calibration:
LED reference module
Signal Detector:
Photomultiplier (PMT) photon counter
Reaction Disk:
150 positions
Temperature: 
Accuracy: 37±0.3°C, Fluctuation: ±0.1°C
Mixing Unit: 
Non-touch vortexer mixing
Cuvette Loading and Waste Collection
Cuvette: 
Disposable plastic cuvettes
Cuvette Loading:
Trays continuous loading and offloading, 126 cuvettes per tray, 
up to 9 trays (1134 cuvettes) can be loaded in one batch
Waste Collection: 
2 waste boxes, overflow detection
Working Conditions:
Power Supply: 
110V/115V~, 60Hz; 220V-240V~, 
50Hz; 220V/230V~, 60Hz
Input Power: 
3200VA
Operating Temperature:
 15°C~30°C
Relative Humidity: 
35%RH~80%RH, without condensation
Dimension:
Main Unit: 1450mm*1020mm*1200mm (W*D*H) 
SDM: 710mm*1020mm*1200mm (W*D*H) 
Total: 2150mm*1020mm*1200mm (W*D*H))
Weight: 
750kg
Altitude Height:
-400m~2000m (optional external compressor for 2000m~4000m)